Finding #1
In the submitted protocol, the laboratory states: "To ensure all manual differentials performed are comparable, . . .manual differentials will be compared annually amongst the competent staff. . ." The laboratory provided no written procedure as to how "manual differentials will be compared annually." In addition, the regulation requires the laboratory to conduct such evaluations twice a year, not annually.

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.

The laboratory failed to address and provide acceptable evidence of correction consisting of: what measure has been put in place or what systemic changes have been made to ensure the deficient practice does not recur; and, how the corrective action(s) is being monitored to ensure the deficient practice does not recur.

Findings #2
Although the laboratory submitted a method/instrument comparison protocol (Ex. A, Tab 28), we were unable to determine if the protocol was effectuated. In its submission, the laboratory states: "The lab's technical supervisors and the quality system director will be responsible for ensuring that these procedures are followed." However, the laboratory's submission did not include any documentation to indicate that the technical supervisors or the quality systems director had been trained on the method/instrumentation comparison protocols.

The laboratory's submission also includes the following statement:

Because TPS's [Theranos Proprietary Devices] were factory calibrated by Theranos, the laboratory did not perform additional method correlation studies. As part of this investigation, Theranos examined the correlation of different TPS to evaluate their concordance. Namely, for SHGB, TT3, Vitamin B12 and Vitamin D, patient distributions were compared across representative TPS's in use during the same period (Exhibit E Tab 12-A; Tab 24-A, Tab 33-A. These data showed similar patient distributions across the different devices. In addition, comparison of assay results across different assay levels and different TPS's demonstrates that the instruments were well correlated to each other (Exhibit E, Tab 12-A; Tab 24-A, Tab 33-A). Namely, the median result for each TPS was within the total allowable error thereby demonstrating that the same assay reference ranges were appropriate for TPS.

The laboratory's response indicates that not all TPS would be included in the correlation studies rather "across representative TPS in use during the same period." The laboratory submitted correlation studies for Sex Hormone Binding Globulin (SHBG) and Vitamin D (VitD) only (Ex E, Tabs 12-A, 24-A and 33-A, respectively). However, these studies did not include a date on either

19