unclear as to how this article applies to this deficient practice, and note that the written procedure submitted by the laboratory at Ex. A, Tab 24 does not incorporate monitoring temperature by MKT.
The laboratory also submitted an article related to the storage of bacterial samples for optimal viability (Ex. N, Tab 22). We note that the submitted written procedure relates to determining acceptable temperature viability of bacterial samples does not include any references to this article. We are unclear as to how this article applied to this deficient practice.
The laboratory submitted no evidence to support the following conclusion for the Sanyo JP Lab 7059 -20 freezer: "[The laboratory] determined that higher storage temperature had no effect on the calibrators or controls stored in this freezer. There is no patient impact from the change in temperature during the time period."
In addition, the laboratory stated for another freezer (1 BUGS 7120 -80 freezer) that the "MKT remained between -70 and -86 C for the entire date range." However, the laboratory indicated that "a small number of bacterial cultures. . . labeled to be stored at -80C or colder" were stored in this freezer. It was unclear as to why it was appropriate to store "bacterial cultures. . . labeled to be stored at -80C or colder" to be stored at temperatures of -70C to -79C.
To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.
The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.
Finding #2
The laboratory submitted a protocol for a Master Validation Plan (Ex. E, Tab 30B) which required the %CV of replicates to be not more than 20% (25% at the lower and upper limits of detection). The submitted protocol was different than the protocol presented to the surveyor at the time of the onsite survey which required the %CV of replicates to be not more than 15% (20% at the lower and upper limits of detection). The %CV in the protocol presented to the surveyor at the time of the onsite survey matched the %CV of the predicate devices. It is unclear as to why the TPS, which is covered by the Master Validation Plan, can now have %CV's for replicates greater than the predicate devices; no explanation was submitted. In addition, we note that the %CV's (18.7% -63.8%) included in the submission does not appear to have been evaluated using this new protocol nor was it addressed in the laboratory's response provided at Ex. E, Tab 1."
22