At the time of the onsite survey, the laboratory maintained no documentation showing that the validation documents for two Siemens Advia 2120i instruments had been reviewed and approved by the laboratory director.
In Ex. G of the submission, we again found no information to indicate the Siemens Advia 2120i validation documents have been reviewed and approved by the director.
Because the laboratory has not shown whether it has corrected this deficient practice, we are also unable to determine whether the laboratory's QA mechanisms can monitor the laboratory's corrective actions and ensure this deficient practice does not recur.
The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.
Finding #6
The submission references "Ex. I, Tabs 2-6." We located these tabs, but found no documentation in Tabs 2, 5 and 6.
See our review of 05413.
Although the submitted protocol (Ex. A, Tab 31, § 8.1.2) included the requirement to review all new manufacturer package inserts and vendor notifications, the laboratory did not provide a response to the citation other than to state: "Theranos did not have the documentation for the MNPT calculation done for lot #539280 in 3/2015 before it was put in use for patient testing." It is unclear what investigation was performed to determine how the Mean Normal Prothrombin Time (MNPT) for Dade Innovin lot number 539280 was defined and entered into the Siemens BCS-XP.
In its submission the laboratory states "training has occurred," however, training documentation provided in Ex. A Tab 32 did not include all testing personnel listed on the CMS-209 (Ex. L, Tab 1).
To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.
The laboratory failed to address and provide acceptable evidence of correction consisting of: what measure has been put in place or what systemic changes have been made to ensure the deficient practice does not recur; and, how the corrective action(s) is being monitored to ensure the deficient practice does not recur.
35