Page:Cms-letter-theranos.pdf/4

From Wikisource
Jump to navigation Jump to search
This page has been validated.

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.

The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.

D2128
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

The submission references "Ex. A, Tab 4, § 7.1.2.2.d.2." We found no such reference contained in the materials provided to CMS.

Although the laboratory's submitted protocol indicates that unsatisfactory PT results would be investigated using CL FRM-00006-F1, the laboratory provided no documentation showing that the required form was used to investigate the unsatisfactory PT result. The laboratory also states that the "lab has investigated the one unacceptable challenge and documented its investigation and conclusions;" however, CL FRM-00006-Fl documenting the investigation was not submitted per the laboratory's protocol.

In the submission, the laboratory concludes:

  • "Not enough patient data available for meaningful analysis"
  • "No evidence of systemic errors"
  • "No patient impact is expected"

However, no information as to how the laboratory came to these conclusions related to patient outcomes was submitted. Documentation contained in Ex O, Tab 3 shows that the laboratory received a passing score of 80% for blood cell identification. While this is true, the CLIA regulations require that all unacceptable results have remedial action taken and that such remedial action be documented. The submission does not contain evidence of remedial action.

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.

4