portion of the second submission (page 82, footnote 14), so our response is based on the remaining 11 testing personnel. The training documents related to the Innovin (thromboplastin) case study include six of the 11 on the Acknowledgement Form provided by the laboratory in Ex. BB, Tab 14D. We are unable to determine if the other five testing personnel have retrained using the case study. We also note that the laboratory submitted training documents (Ex. BB, Tab 14B) on the revised reagent qualification procedure (CL SOP 07010) in Ex. BB, Tab 14C, which was missing documentation for one individual. We also observed that a change was made to the procedure (CL SOP 07010, Rev. D) which was approved on March 25, 2016. However, training for six of the testing personnel was completed between February 1 and 4, 2016, so they have not been trained on the revised procedure.
Additionally, in its second submission, the laboratory submitted a revised quality assessment mechanism, but submitted no documentation to indicate that the revised quality assessment mechanism had been effectuated. In addition, the laboratory stated that "training on the revised QMPI procedures has occurred. (Ex. BB, Tab 7 B-C)." We note that Ex. BB, Tab 7 B-C contains incomplete training documentation (one of training records did not include the same documents consistent with the other two records) for only three laboratory personnel, none of whom appear on the laboratory's Form CMS-209 dated February 8, 2016. It is also unclear to CMS if the laboratory's new QMPI will be effective as the above issue (i.e., disparity of MNPT values for Innovin lot number 539280) has not been recognized by the laboratory in either submission.
The laboratory again failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information as set forth in our January 25, 2016 and March 18, 2016 letters. In addition, the laboratory again failed to address and provide acceptable evidence of correction consisting of: what measure has been put in place, what systemic changes have been made to ensure the deficient practice does not recur, or how the corrective action is being monitored to ensure the deficient practice does not recur.
D6093
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.
Finding #1
See our review of D5481. For D5469, see our Quality Assessment review under the General Comments.
Finding #2
See our review of D5481.
We note that in the second submission, the laboratory states that it "issued corrected reports voiding PT/INR test results reported for the period of October 2014 through September 2015. (Ex. AA, Tab 7). The laboratory stopped using the BCS XP instrument on September 17, 2015, and has not used it since then." Documentation given to the surveyor by the laboratory at the time of the onsite survey revealed that the laboratory did, in fact, use the BCS XP for patient testing after September 17, 2015.
25