The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

See our review of D6115.

D6115
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

In the April 1, 2016 portion of the second submission on page 84, the laboratory stated that the TPS 3.5 was "fully retired in early-August 2015." However, based on documentation supplied by the laboratory at the time of the onsite survey, the TPS was not used for patient testing after June 25, 2015. We note that no explanation was submitted regarding the disparity between the end date provided at the time of the onsite survey and the end date provided in the second submission. The laboratory's second submission does not provide an explanation for the issues related to corrected results, accuracy, %CV, precision study, reference range,% recovery, and allowable bias.

Although the term "medical decision level" (MDL) is now defined by the laboratory, we find no documentation to indicate which analytes have a medical decision level, how that medical decision level is determined, and what the distribution of specimens would be if an analyte does not have a medical decision level. It is also unclear to CMS if patient results are reported with an MDL, how that will impact the normal and abnormal ranges for a given analyte, especially when the MDL does not overlap with or excludes the normal range.

Since the laboratory ceased patient testing using the TPS 3.5 and voided all patient test results performed using the TPS from 2014 and 2015, CMS was unable to determine if the updated method verification procedure had been effectuated.

The laboratory again failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information as set forth in our January 25, 2016 and March 18, 2016 letters. In addition, the laboratory again failed to address and provide acceptable evidence of correction consisting of: what measure has been put in place, what systemic changes have been made to ensure the deficient practice does not recur, or how the corrective action is being monitored to ensure the deficient practice does not recur.

D6124
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

The "Training and Competency, Revision C" procedure submitted in Ex. BB, Tab 9 has an approval date of March 24, 2016. Revision C includes the competency assessment requirement for direct observation of routine patient testing and performance of instrument maintenance and function checks as well as the six CLIA regulatory procedural requirements set forth at 42 C.F.R. § 493.1451(b)(8). It is unclear as to how training occurred prior to March 24, 2016 when the updated procedure, which included changes to the competency assessment procedure related to