Theranos provided CMS five passcode protected flash drives with its second submission labeled Exhibits KK, MM, NN, 00, and PP. After careful review, we note the following related to these flash drives:

  • The information on the flash drives was difficult to evaluate and access. Corrected patient test reports and notification of corrected patient test reports for any specific patient specimen accession number were spread over several flash drives, and the corrected patient test reports were not always associated directly with the confirmation of notifications and receipts. This required all flash drives be reviewed in their entirety for any specific accession number to determine whether complete records were received. We were unable to search the flash drives by any criteria (e.g., patient name, date of visit, etc.) other than by the specimen accession numbers provided in the binder labeled Ex. MM.
  • We are uncertain as to what the laboratory intended to submit as a "complete" record for each corrected patient test report submitted. Some of the corrected patient test reports had associated facsimile coversheets, and other corrected patient test reports had associated facsimile receipts, but not both. In other cases, we found only facsimile coversheets or receipts, but no associated correct patient report. We are unclear as to the following:
    • For those specimen accession numbers that had only facsimile coversheets, how did the laboratory verify receipt of the corrected patient reports?
    • For those specimen accession numbers that had only facsimile receipts, how does the laboratory (and CMS) know what information was actually transmitted to the authorized person?
    • How were the receipt confirmations generated?

It appears the information provided on the flash drives did not include "complete" documentation for all specimen accession numbers listed in the binder labeled Ex. MM. Furthermore, some of the specimen accession numbers listed in binder Ex. MM could not be found on any of the flash drives.

Consequently, we could not determine whether the laboratory provided documented evidence showing what corrective actions were taken for all patients found to have been affected by the deficient practice, and what corrective actions were taken for any other patients identified as having the potential to be affected by the same deficient practice.

  • Quality Assessment

Although related to all deficiencies cited, this general comment specifically addresses the laboratory's second submission responding to the deficiencies cited under D2094, D2128, D5217, D5311, D5391, D5393, D5403, D5413, D5421, D5423, D5429, D5437, D5447,